Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
Inspection and maintenance 55
Page 55 of 64
Medical Solutions
• If the deviation is larger than 1.7%, correct the coincidence of the light and radiation
fields by adjusting the nuts marked "D” and "E”. (Fig. 30 / p. 51)
X
• Check the setting by releasing another exposure.
• If necessary, repeat the adjustment until the deviation of the light and radiation fields
lies within the acceptable tolerance.
QSQ Deviation ((A + C) / SID)
QSQ Deviation ((B + D) / SID)