2.16 Checking the manual exposure interrupt
The user must be able to interrupt the exposure at any time.
• Set 55 kV, 160 mAs for MOBILETT Plus/Plus M.
• Set 55 kV, 80 mAs for MOBILETT Plus E.
X
• The message can be erased by pressing the ”kV+” button.
SIE Message "USE20"
SIE 10 short beep signals
NOTE
All covers must be remounted.
• Release an exposure and immediately interrupt it by letting go of the exposure switch.
The message ”USE 20” appears on the display indicating that the exposure release
switch was interrupted during the exposure. In addition, 10 short "beep" signals
sound.