26 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 26 of 64
Medical Solutions
Fig. 3:
SIM Brake lining
• Check the brake lining (Fig.3/p.26). If the lining is worn down to less than 1 mm, it
must be replaced. Refer to the section on ”Adjustment of the hand brake/parking brake”
in the service instructions.
• Remount the wheel.
SIM Secure mounting
SIM Ease of movement
• Check that the wheel is securely mounted and that it turns freely.
• Check the other wheel following the same procedure.
2.2.2 Front wheels
Fig. 4: Fig. 5: