28 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 28 of 64
Medical Solutions
2.2.4 Turn plate (MOBILETT Plus/Plus M only)
Fig. 7:
PMF Button illumination
• Press one of the turn plate buttons. The buttons must light up when the turn plate is acti-
vated.
PMF Function
• Check whether the system can be rotated around the turn plate (Fig.7/p.28).
• Press the turn plate button again and check whether the turn plate retracts to the park-
ing position.
• Repeat the same test with the other turn plate button. If a problem arises, proceed ac-
cording to the section ”Adjustment and replacement of the turn plate” in the service in-
structions.
• Connect the mains cable and switch the system ON.
• Move the unit to a level surface.