30 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 30 of 64
Medical Solutions
2.2.6 Transport safeguard
SIM Damage
SIM Locking
Check the transport safeguard (Fig.9/p.30) for mechanical damage. In case of dam-
age, the transport safeguard must be replaced.
Check that the handle of the arm system locks securely into the transport safeguard.
2.2.7 Handles
SIM Mounting
Check that the handle of the support arm is securely mounted. If necessary, tighten the
screws with a torque of 15-20 Nm.
2.2.8 Collimator adjustment knobs
SIM Mounting
Check whether the adjustment knobs on the collimator are tightened securely
(Fig. 10 / p. 30).
If necessary, tighten both Allen screws on each of the knobs.
Fig. 9:
Pos. 1 Transport safeguard
Fig. 10: