Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
Inspection and maintenance 29
Page 29 of 64
Medical Solutions
2.2.5 Brakes
Fig. 8:
WARNING
There is a risk of life-threatening electric shock.
¹ Make sure protective measures have been taken prior to
opening the unit.
¹ Disconnect the system from the mains.
¹ (Protective measures when working on the capacitor
bank/p.14).
SIM Uniformity
SIM Braking force
SIM Locking
• Check whether the hand brake is sufficient to stop the unit and that it does not pull to the
left or the right when the brake is applied.
• Lock the brake.
¹ It should not be possible to push the MOBILETT using average force. Slight
skidding is permissible when applying a force greater than 350 N.
• Remove the front, rear, left and right covers.
• The brake force can be increased by increasing the tension on the brake cable
(Fig.8/p.29). Additional information can be found in the section on ”Adjustment of the
hand brake/parking brake” in the service instructions.