Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
Inspection and maintenance 39
Page 39 of 64
Medical Solutions
Pressing the ”kV+” button calls up the next test program. As soon as the last test program
has been reached and ”kV+” is pressed again, the display begins again with the program
version.
The "kV-" button allows you to run the test programs in reverse order. The "collimator
lamp" button is used for the ON/OFF function, i.e., it starts or stops the displayed test pro-
gram.
Example of the display for P06 after activation of the collimator lamp (starts the formation
of the capacitor bank):
The test program is stopped by pressing the "collimator lamp" button a second time.
Example of the display for P06:
Program P06 is now stopped. A new service program can be selected with the ”kV+/-”
buttons or program no. 6 can be restarted by pressing the "collimator lamp" button.
NOTE
Messages appearing in service mode, with the exception of "Quick
adjustment", are only displayed, i.e. they are not incorporated in
the error memory.