EasyManua.ls Logo

Siemens MOBILETT Plus - Page 43

Siemens MOBILETT Plus
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
Inspection and maintenance 43
Page 43 of 64
Medical Solutions
X Set the following exposure parameters and release an exposure:
* The measurement inaccuracy of the respective measurement device must be sub-
tracted from these limit values.
Setting
Proceed according to section "P10 - Fast adaptation" in the service instructions in order
to adjust excessively high deviations.
PMF 52 kV, 50 mAs
PMF 81 kV, 20 mAs
PMF 133 kV, 20 mAs
PMF 125 kV, 20 mAs
kV meter
selection
Exposure parameter Limit values* to be maintained
DC voltage
52 kV 50 mAs 49.0-55.0 kV
81 kV 20 mAs 77.0-85.0 kV
MOBILETT Plus/Plus M MOBILETT Plus/Plus M
133 kV 20 mAs 126.4-139.6 kV
MOBILETT Plus E MOBILETT Plus E
125 kV 20 mAs 118.8-131.2 kV

Other manuals for Siemens MOBILETT Plus

Related product manuals