Revision Date: 3/13/2019
20
Coil Warnings - Protocol Was Fixed
If you see warning messages while scanning, please read the message and address the
issue. DO NOT just click “OK”. Ask a tech for help if you are uncomfortable with
resolving the issue on your own. Please be vigilant, as we do not refund scans for this
type of issue.
Main Reasons for Coil Warnings
1. The coil wasn’t plugged at all or properly
2. The subject was registered and the protocol was loaded into the exam card before
plugging in the coil
Anterior or Posterior Protocol Fixed Warning Messages
1. The pop up message will let you know that it is turning off
the elements that are not plugged in properly
2. If you click “OK”, you will be approving this change and
you will lose signal from that part of the coil
3. Data collected in the manner will be suboptimal and most
likely unusable
Resolution
1. Check the coil plugs
2. Reseat the plug if necessary
3. Make sure the LCD screen displays the anterior and
posterior elements
4. Close your subject
5. Re-register the subject
6. Try again
Wrong Coil Protocol Fixed Warning Messages
1. The wrong coil was put on the scanner bed
The message will let you know that it is changing
the coil to the one you put on the scanner bed
Your data is not be “bad” per se, but you won’t be
using the proper coil for your study
Resolution: change the coil, close and re-register
your subject
2. The protocol was not setup for the correct coil
The message will let you know that it is changing
the coil to the one you put on the scanner bed
Resolution: this is the only scenario where it is ok to click “OK” and proceed
Let a technologist know that your protocol is setup with the wrong coil so it can
be fixed
Anterior (HC1,3,5,)
portion of 64ch coil
not active
Anterior (HEA)
portion of 32ch
coil not active
Coil changed
from 64ch to
32ch