EasyManuals Logo

Siemens Prisma 3T User Manual

Siemens Prisma 3T
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
Revision Date: 3/13/2019
27
Preparation Scans related to Regular BOLD Sequences
1. To ensure that a steady-state magnetization is achieved when multiple averages or
measurements are acquired, a number of preparation scans are performed at the start of
the measurement
2. The number of preparation scans is set within the sequences, and chosen so that the
preparation time is longer than 3 seconds
TR = 500 ms 7 preparation scans
TR > 501 ms 6 preparation scans
TR > 601 ms 5 preparation scans
TR > 751 ms 4 preparation scans
TR > 1001 ms 3 preparation scans
TR > 1501 ms 2 preparation scans
TR = 2000 ms 2 preparation scans
TR = 2500 ms 2 preparation scans
TR = 3000 ms 2 preparation scans
TR > 3001 ms 1 preparation scan
3. The preparation scans result in a longer scan time for the first measurement in a time-
series acquisition, which should be taken into account when performing BOLD imaging
4. The sequence does not send external trigger pulses when performing these preparation
scans, so that they do not have to be considered when triggering external devices for
stimulus presentation in BOLD imaging studies
5. This formula does not apply to multiband bold sequences
6. Turning iPAT on will add 1 additional prep scan
7. Turning introduction “on” will add 1 additional prep scan

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Prisma 3T and is the answer not in the manual?

Siemens Prisma 3T Specifications

General IconGeneral
BrandSiemens
ModelPrisma 3T
CategoryMedical Equipment
LanguageEnglish

Related product manuals