Revision Date: 3/13/2019
25
Participant Registration Errors
1. You cannot confirm your registration due to the below error
This happens when there are still active sequences in the Exam Card
Go back to the Exam card and stop (red square) or Skip (>) the remaining
sequences in the queue
This happens when a participant aborts a scan early or the group runs out of time
and stops scanning before their protocol is complete
Also note, you will not be able to close the patient on the Exam card until the
sequences are stopped, skipped and/or deleted
2. The registration exam button is grayed out so you cannot click it
This means you have forgotten to complete a required field
Required fields are in bold
NOTE: Date of Birth is bold, but we use Age instead please do not input a
DOB once you input the age a general DOB will be generated
NOTE: Referring Physician is not bold but is required to access your data