SITRANS FM100
A5E48762204-AA page 59
Index
[hex]
Name
Description
Factory default
Max Value
Min Value
Length
[Bytes]
Data Type
Access
0x0110
OUT1Source
Source for the
output
(production
setting)
0
(0) - Flow
(1) - Volume
(2) - Temperature
(3) - Part Volume
1
UIntegerT
0x0111
OUT1Type
Configuration of
the output: 0-
20mA, Pulse,
Frequency
(production
setting)
8
(0) - disabled
(1) - Alarm Output
(2) - 4-20mA
(3) - 0-20mA
(4) - 2-10V
(5) - 0-10V
(6) - Pulse Output
(7) - Frequency Output
(8) - M12COM
(9) - IO-Link
(10) - Control Input
1
UIntegerT
0x0112
OUT1Alarm
Function
Limit or window
function for the
alarm output
0
(0) - Limit
(1) - Window
1
UIntegerT
R/W
0x0113
OUT1AlarmOutput
Type
Alarmoutput NPN,
PNP or Pushpull
0
(0) - NPN
(1) - PNP
(2) - PushPull
1
UIntegerT
R/W
0x0114
OUT1AlarmSwitch
Function
Alarmoutput
normally opened
or closed
0
(0) - normally opened
(1) - normally closed
1
UIntegerT
R/W
0x0115
OUT1Alarm
Threshold
Threshold for the
alarmoutput
1,0
MRE
MRS
4
FloatT
R/W
0x0119
OUT1AlarmLower
Threshold
Threshold for the
alarmoutput used
by the
windowfunction
1,0
OUT1AlarmT
hreshold
MRS
4
FloatT
R/W
0x011D
OUT1Alarm
Hysteresis
Switching
hysteresis for the
alarmoutput
1,0
9999,0
-9999,0
4
FloatT
R/W
0x0121
OUT1Alarm
SuppressionFactor
How many times
the threshold must
be hit in order to
switch the alarm
output
0
60
0
1
UIntegerT
R/W
0x0122
OUT1Alarm
Suppression
Direction
for which direction
the suppression
factor is used
0
(0) - Up
(1) - Down
(2) - Both
1
UIntegerT
R/W