Commissioning
6.4 Commissioning via PDM
SITRANS FUS080 transmitter
38 Operating Instructions, 07/2012, A5E03059912-02
3. Depending on the SONOKIT type (1-path or 2-path), enter additional sensor data:
Parameter 1-path 2-path
Number of paths 1 2
Outer pipe diameter D
U
D
U
Pipe wall thickness t t
Average path distance* H H
Path 1 angle A1 A1
Path 2 angle - A2
Average transducer 1 distance L1 L1
Average transducer 2 distance - L2
Velocity (auto calculated with Qmax) V V
* For crossed paths solutions with no path displacement: Enter "0".
Figure 6-2 1-path sensor data