Parameter lists
A.3 Diagnostic
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
73
Modbus
Address
Parameter Default
value
Value range Access
level
Description
40780 163 Previous settling date 01-01-
2000
Read Previous settling date where value
of totalizer 1 was stored
40783 164 Previous totalizer 1 value 0 dd-mm-yyyy Read Previously stored value of totalizer 1
A.3 Diagnostic
Table A- 3 Diagnostic
Modbus
Address
Parameter Default
value
Value
range
Access
level
Description
Diagnostic
40804 500 Latest service date 01-01-
2000
dd-mm-
yyyy
R/W* Latest service date (can also be used
for date of installation)
40808 501 Operating hours since
power up
0 3.4E+38 h Read Total operation hours since last power
up
40807 503 Numbers of power up 0 0 ... 65535 Read Total number of power-ups since first
power-up.
Alarm
43005 200 Fault status One or
more of
the
following
values: 1,
2, 3, 4, ...
up to 16
Read Fault status
1: Path 1 error
2: Path 2 error
3: Dual slope calibration error
4: Database checksum error
5: Low power warning
6: Flow overload warning
7: Pulse A overload warning
8: Pulse B overload warning
9: Consumption interval error
10: Not used
11: Not used
12: Not used
13: Not used
14: Not used
15: Not used
16: Not used
40346 202 Date of fault log reset Last reset
date
dd-mm-
yyyy,
hh:mm:ss
Read Date of last fault log reset
14 204 Reset the fault log and
faults
No No, Yes R/W Reset the fault log and faults
15 205 Call up acknowledge No No, Yes R/W Select 'Yes' to reset active call-up