Parameter lists
A.2 Output
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
71
A.2 Output
Table A- 2 Output parameters
Modbus
Address
Parameter Default
value
Value range Access
level
Description
43023 100 Actual date and time Product
dependent
R/W Actual date and time (day-month-
year and hours:minutes:seconds)
43006 101 Totalizer 1 0 -2000000001
...
+2000000001
Read Volume for totalization register 1
43010 102 Totalizer 2 0 -2000000001
...
+2000000001
Read Volume for totalization register 2
43014 103 Customer totalizer 3 0 Read Customer totalizer 3 based on
totalizer 1 set up
19 104 Reset customer totalizer 3 No No, Yes R/W Reset of customer totalizer
40732 105 Customer totalizer 3 reset
date
last reset
date
dd-mm-yyyy,
hh:mm:ss
Date and time when customer
totalizer has been reset
43001 106 Flow rate Read Actual flow value
43136 107 Relative flow rate 0 Read Actual flow value in relation to Qmax
43035 108 Actual velocity for path 1 0 Read The actual flow velocity for path 1 in
m/s (SI-unit)
43046 109 Actual velocity for path 2 0 The actual flow velocity for path 2 in
m/s (SI-unit).
Valid for 2-path solutions only
43026 110 Path 1 gain step 15 1 to 15 Read Amplifier setting when measuring
path 1
43037 111 Path 2 gain step 15 1 to 15 Read Amplifier setting when measuring
path 2.
Valid for 2-path solutions only
Pulse A B
40817 430 Output A enable Yes No, Yes R/W Select 'Yes' to enable output A
40818 431 Pulse A function Forward Forward,
Reverse,
Forward net,
Reverse net
R/W Pulse output calculation on pure
forward flow or pure reverse flow -
or net flow for forward and reverse
pulse flow.
Valid if pulse output A is enabled
40819 432 Amount per pulse A 1.000000
(m
3
)
0.000001 ...
1000000000
R/W Volume per pulse and selected unit
Valid if pulse output A is enabled.
40821 433 Pulse width for pulse A 5 ms 5, 10, 20, 50,
100, 200, 500
ms
R/W Pulse length when the pulse is
active.
Valid if pulse output A is enabled
40822 440 Output B enable Yes No, Yes R/W Select 'Yes' to enable output B
40823 441 Pulse B function Alarm Pulse, Alarm,
Call up
R/W Configuration of output B as pulse -
alarm or call up function.
Valid if pulse output B is enabled