Service and maintenance
8.2 Battery replacement
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
51
3. Connect battery.
Ensure that wire is inserted into small channel leading from plug to battery.
4. Remount battery cover.
Note
Every time a
battery is fitted and connected, the unit runs a start-up routine.
A battery replacement does not influence accumulated process values. Accumulated
values can only be reset using SIMATIC PDM.