Standardizing the measuring system
Before starting the apex measurement, a functional check or
standardization of the measuring system can be performed by shorting
the electrodes. This cancels out any inaccuracies caused by skips in
impedance in the measurement setup.
➢ Short the electrical measurement system. Plug in the files and hold
them directly against the mucosal electrode.
If a short signal is heard and the distance display appears with no
bars, the standardization was successful. If not, check the
electrical connections for signs of damage.