EasyManua.ls Logo

Soredex CRANEX Novus e - Page 60

Soredex CRANEX Novus e
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. Quality Assurance CRANEX Novus e
54 User manual 208898
NOTE:
When using the 0.8 mm copper lter the
expected Line Pair resolution achievable with
the unit is 3.1 LP/mm or better, with all four low
contrast resolution holes distinguishable.
The line pair and low contrast resolutions
actually distinguishable from the image
depend, not only on the unit capabilities, but
also on other parts of the system, for example
the dental imaging software and the display
monitor being used.
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4
Copyright © 2015 by PaloDEx Group Oy. All rights reserved.

Table of Contents

Other manuals for Soredex CRANEX Novus e

Related product manuals