12. Check of cut module output power at a test resistor
12.1 Limits: min. 6 max. 14
Power selection 1; Cut module defective
12.2 Limits: min. 60 max. 80
Power selection 2; Cut module defective
12.3 Limits: min. 110 max. 150
Power selection 3; Cut module defective
12.4 Limits: min. 160 max. 220
Power selection 4; Cut module defective
12.5 Limits: min. 220 max. 280
Power selection 5; Cut module defective
12.6 Limits: min. 20 max. 47
Test current limiting; Cut module defective
12.7 Limits: min. 120 max. 190
Test current limiting; Cut module defective
13. Check of coag module output power at a test resistor
13.1 Limits: min. 15 max. 25
Power selection 1; Coag module defective
13.2 Limits: min. 35 max. 55
Power selection 2; Coag module defective
13.3 Limits: min. 60 max. 80
Power selection 3; Coag module defective
13.4 Limits: min. 80 max. 110
Power selection 4; Coag module defective
13.5 Limits: min. 100 max.140
Power selection 5; Coag module defective
Power-supply data
Mains connection: 230v ± 10%, 50 Hz± 5%
internally switchable to
115v ± 10%, 60 Hz ± 5%
Mains fuses: built into mains-filter containing 2 fuses T 4A/250V
according to DIN 41 662 for 230V mains-voltage
respectively T 8A/250V for 115v mains-voltage
Mains cable: length 2.5 m with protective conductor, mains-part according
to DIN 49 441 resp. CEE 7 p.VII, equipment-part according
to DIN 49 457 p.1., Plug and socket vulcanized to cable.
The cable must comply with actual standards and must have
a cross sectional area of 1 mm
2
Potential equalization: Connector according to DIN
Power consumption: 1000 VA