EasyManua.ls Logo

STERO SG - SOAP PROGRAMMING (7627); EDIT PARAMETERS

STERO SG
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SOAP PROGRAMMING (7627)
Service Programming
EDIT PARAMETERS
PRIME PUMPS
EXIT MENU
1. Press MENU button on keypad.
2. Enter SERVICE MENU.
3. Enter Service code 7627.
EDIT PARAMETERS
Parameter
Name
Cold Units
(Stero SU-L)
(Stero Glass
Washer SG)
Hot Units
(Stero SU-H)
Description Possible Values Default Values
Detergent
Pump
Stero SU-L
Stero SG
Stero SU-H
Detergent pump
option.
Disabled / Enabled Enabled
Detergent
Flow Rate
(Amount)
Stero SU-L
Stero SG
Stero SU-H
Detergent flow rate
(amount) measured
in mL per cycle.
OFF through 20.0 mL 7.9 mL
Rinse Aid
Pump
Stero SU-L
Stero SG
Stero SU-H
Rinse Aid pump
operation.
Disabled / Enabled Enabled
Rinse Aid
Flow Rate
(Amount)
Stero SU-L
Stero SG
Stero SU-H
Rinse Aid flow rate
(amount) measured
in mL per cycle.
OFF through 8.8 mL 1.8 mL
Sanitizer
Flow Rate
(Amount)
Stero SU-L
Stero SG
N/A
Sanitizer flow rate
(amount) measured
in mL per cycle.
Off to 100 mL 60 mL
Detergent
Threshold
Stero SU-L
Stero SG
Stero SU-H
Sets detection
amount.
Sets maximum
allowable chemical
sense reading (mV)
for detergent.
0-4000 (by 100's) 1100
Rinse Aid
Threshold
Stero SU-L
Stero SG
Stero SU-H
Sets detection
amount.
Sets maximum
allowable chemical
sense reading (mV)
for rinse aid.
0-4000 (by 100's) 2500
SG, SU-L, SU-H DISHWASHERS - SERVICE PROCEDURES AND ADJUSTMENTS
Page 25 of 59 F45577 (1215)