EasyManua.ls Logo

Stiegelmeyer Deka - Inspection by the User

Stiegelmeyer Deka
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
Inspection by the User
114
Instruction Manual
18.4 Inspection by the User
Inspection
Ok
Not OK
Description
of Defect
What to Check...
Check for...
Visual Inspection of the Electrical Components (if installed)
Handset: Housing + key-
pad + display + cable:
No damage
Handset: Cable + cable
suspension
Securely suspended + routed
without risk of being crushed in
the bed frame
Mains cable
No damage, safe routing
Control box / locking box:
Housing + keypad + cable
No damage, routed without
risk of being crushed in the
bed frame
Visual Inspection of the Mechanical Components
Patient lifting pole, location
sleeves
Damage, deformation
Grab handle with strap
No damage
Chassis
Damage, deformation
Mattress base, covers
Damage, deformation
Performance Check of the Electrical Components (if installed)
Handset / locking box
Performance check according
to
Electric Actuator System » 58
Performance Check of the Mechanical Components (if installed)
Castors
Safe braking according to
Moving and Immobilising the
Bed » 49
Emergency lowering of
backrest (CPR)
Lowers when release lever is
activated in accordance with
Lowering the Backrest by
Hand » 70
Side rails, safety sides
Performance test, securely
engaged in accordance with ¾
Safety Sides (Swivelling) » 66
Accessories (e.g. patient
lifting pole, grab handle
with strap, side rails)
Securely fixed, undamaged;
suitability according to ¾ Safe-
ty Sides (Swivelling) » 66
Inspector’s signature: Inspection results:
Date:

Table of Contents