15.2 Troubleshooting for Electric Beds
The following table is a guide to rectifying faults:
Problem
Possible Causes
Solution
Manual adjustment of backrest
is not possible
Rod that connects release
lever and gas spring is bent or
has become detached
Readjust at the release lever
or secure them
Manual adjustment of thigh rest
is not possible
Rod that connects release
lever and gas spring is bent or
has become detached
Readjust at the release lever
or secure it in place
Manual tilting of mattress base
is not possible
Bowden cable too long / has
become detached
Adjust length of Bowden cable
Bowden cable jammed /
kinked
Rectify the cause; fix securely;
replace if necessary
Manual tilting of mattress base
cannot be locked
Bowden cable too short
Adjust length of Bowden cable
Bowden cable jammed / does
not reset itself
Rectify the cause; fix securely;
replace if necessary