Use in an Emergency
Emergency: Setting CPR Position
Deka
69
10 Use in an Emergency
10.1 Emergency: Setting CPR Position
The CPR position allows all parts of the mattress base to be lowered quickly, particularly for re-
suscitation purposes. It is also suitable for moving the bed to a specific low position for the patient
to sleep at night (fall prevention).
The adjustment position is preset in the factory:
x Backrest and thigh rest horizontal, simultaneously
x mattress base in lowest position
WARNING
Risk of Injury
This function is only intended for medical emergencies and not for everyday use!
This function demands the user’s utmost attention since various adjustments are activated simul-
taneously.
If this warning is disregarded, patients and other people could be exposed to danger and
property could be damaged.
The CPR position can also be activated even if the corresponding adjustment functions are
locked.
10.1.1 On the Handset
The CPR position can only be activated
when the staff control level is set (see
Handset with Magnetic Locking Function »
58).
1
Press the CPR button on the handset
and keep this pressed.
The backrest and thigh rest are moved
to a horizontal position and the mat-
tress base is also adjusted to its low-
est height.
!
M