Troubleshooting
Troubleshooting for Electric Beds
Deka
95
15.1.2 Fault Indicator (Version EM2 and EM3 Only)
For safety reasons, if a serious error is detected, the functions concerned are electronically locked
out. At the same time, the various errors are displayed on the handset and/or locking box by the
lock indicator flashing differently. This makes it easier to rectify faults.
Once the fault has been rectified, the control unit must be RESET to ensure that the system will
work properly again. The necessary steps are explained in the Chapter 15.1.3 Unlocking the Con-
trol Unit Reset.
15.1.2.1 Handset with Magnetic Locking Function
If the bed is equipped with a handset with a magnetic locking function, the error codes will be
shown by the LEDs on the handset itself.
Possible Fault
Handset fault
An internal fault in the control
unit
Thigh rest motor (socket
5) limit switch
Solution
Replace handset with a new
one
Reset the control unit
Check cable + plug
Check/replace motor
Faults are
shown through
flashing LEDs