34
Instruction Manual
6.6.6 Operating Noise
The operating noise of an electrically adjustable bed is not more than 47 dB (A).
6.6.7 Ambient Conditions
The following ambient conditions must be maintained:
Ambient Conditions for Storage
Minimum
Maximum
Storage temperature
-10°C
+50°C
Relative humidity
20 %
90 % (at 30°C; non-condensing. At
altitude ≤ 2000m)
Ambient operating conditions
Minimum
Maximum
Ambient temperature
+5°C
+40°C
Relative humidity
20 %
90 % (at 30°C; non-condensing. At
altitude ≤ 2000m)
Air pressure
800 hPa
1060 hPa
6.6.8 Electrical Data
Mains Cable: (coiled, anti-kink, with strain relief)
Type
H05 BQ-F 3 x 1 mm
2
(EPR quality)
Handset (EM1 - EM3)
Type
LINAK HB 7x; HB 8x (no locking function)
LINAK HD 8xx (with integrated locking func-
tion)
Operating voltage
24 V DC
Type of enclosure
IP X6, suitable for automatic washing systems (optional)
Locking Box (EM2 - EM3) (Optional)
Type
LINAK ACC
Operating voltage
24 V DC
Type of enclosure
IP X6, suitable for automatic washing systems (optional)
Control Unit (with integrated rechargeable lead-acid battery)
Types
LINAK CB12L (EM1)
LINAK CB16XX (EM2 - EM3); battery optional
Input voltage
AC 230 V, ± 10 %, 50/60 Hz
Current input
max. 2 A (230V)
Output voltage
24 V DC
Output current
Max. 10 A (electronic monitoring and cut-out)
Duty cycle
Intermittent duty: 2 min ON / 18 min OFF
Classification
Protection class I, EN60601-1-compliant type B device with internal power sup-
ply (if fitted with lead-acid batteries), not for use in explosive atmospheres
Protection category
IP X6, suitable for automatic washing systems (optional)
Lead-Acid Battery (Optional)