Do not use scouring agents, abrasive cleaning agents or scouring pads. These substances
can damage the surfaces.
Do not use organic solvents such as halogenated/aromatic hydrocarbons and ketones.
Do not use acidic cleaning solvents.
It is essential that the manufacturer’s dosage advice is followed to prevent damaging
the plastic and metal surfaces!
It is not permitted to clean the bed using a manually operated steel jet nozzle which is, for
example, connected to a steam cleaner/ high pressure cleaner. A minimum distance of 30
cm from the electrical components cannot be guaranteed in this case.
Ensure that no liquid residues remain on any parts of the bed after cleaning or disinfection.
Otherwise the surfaces in these areas may become damaged in the long term.
Despite the excellent mechanical resistance, scratches, knocks, etc., which permeate the
entire coating should be resealed using a suitable repair medium to prevent the penetration
of moisture.
12.3 Cleaning and Disinfection Agents
x For disinfection by wiping, most cleaning and disinfection agents commonly used in hospitals or
care facilities, such as cold and hot water, detergents, alkaline solutions and alcohols, can be
used.
x These agents must not contain any substances that could change the surface structure or the
adhesive properties of the plastic materials.
x The choice of cleaning agents and disinfectants available on the market may change from time
to time. Stiegelmeyer therefore routinely tests the most commonly used materials for compati-
bility.
The most up-to-date list of cleaning agents and disinfectants can be obtained from our service
centre on request: Contact info: » 13.1