EasyManua.ls Logo

Stiegelmeyer Deka - Inspection and Compliance Documents

Stiegelmeyer Deka
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
Inspection Report
110
Instruction Manual
18.3 Inspection Report
Inspection Report following an Inspection of Electromedical Equipment according to DIN
EN 62353 (VDE 0751-1): 2015-10
Customer / med. facility / practice:
Address:
Carried out:
[ ] Repeat inspection
[ ] Inspection prior to initial
operation (reference value)
[ ] Inspection following repairs/servicing
Type of device: [X] Hospital
bed
[  ] Care bed
Protection class:
[X] I
[ ] II
Bed type:
Inventory number:
Location:
Serial number:
Manufacturer: Stiegelmeyer GmbH & Co.
User-specific parts: none
Testing equipment used (type/inventory no.):
1.
2.
MP-RL 93/42 Classification: Class I type B
I. Visual Inspection
Ok
Not ok
Description of Defect
What to Check...
Check for...
Visual Inspection of the Electrical Components (if installed)
Control unit / power
pack:
Type plate: Available,
legible
Housing: Securely fixed,
No cracks / damage
All plugs inserted and locked
in securely?
Motors:
Housing and lifting tubes:
No cracks / damage
/ deformation?
Handset
Housing + keypad + display:
No damage?
Internal cabling: Motor
cable, handset cable,
mains cable, additional
component cables
No damage, securely fixed,
safe routing of cable without
risk of it being crushed when
bed is moved, mains cable
holder available

Table of Contents