Appendix
Inspection Report
Deka
111
I. Visual Inspection
Ok
Not
Ok
Description of
Defect
What to Check...
Check for...
Visual Inspection of the Mechanical Components (if installed)
Type plate + warning labels on
bed
Available on bed frame,
legible
Patient lifting pole, adaptor
sleeves, grab handle with strap
No damage or deformation
Manufacturer's recommendation: Replace grab handle after 5
years
Bed frame: Mattress base,
chassis
No damage, deformation,
no split welded seams
Castors
No damage
Mattress base
No damage or deformation
Safety sides
No damage, cracks or de-
formation
Connecting elements
(screws, bolts, nuts, safety
caps)
Fixed position, complete-
ness
Wearing parts,
Movable joints
No damage or severe wear
II. Electrical Measurement according to DIN EN 62353 (VDE 0751-1): 2015-10
Threshold
Measured
value
Resistance of earth wire
Measuring point: PE pin on
the bed, head end.
Measuring
current
≥2 A
0.3 Ω
Ω
Leakage current of device,
direct/ difference (place bed
with conductive castors in a
way that it is insulated).
0.5 mA
mA
1
Plug the bed mains cable
in the test socket on the
measuring instrument.
2
Connect the measuring
instrument probe to the
PE connector
(mattress base,
head end).
3
For the duration of the
measurements, activate
the motors using the
handset.