Performance Inspection of the Electrical Components (if installed)
Battery powered; capacity
of battery (optional equip-
ment)
Requirements: Battery is
charged + bed is discon-
nected from power supply:
Test: Load bed with approx.
80 kg (=1 person); min. 2
cycles of height adjustment
Up/Down must be possible
before cutting out.
End of travel cut-out for
motors
Automatic cut-out in both
end positions
Handset, operational con-
trols, locking functions
Test according to instruc-
tion manual. No ‘rattling’
when shaken
No abnormal noise level,
no uneven running
Backrest motor and manual
emergency lowering
Control unit / transformer
and motors
Test according to instruc-
tion manual
Strain relief of mains cable
Mains cable firmly fastened
Performance Inspection of the Mechanical Components (if installed)
Lower leg rest (Rastomat)
Engaged, evenly on both
sides
Test according to instruc-
tion manual
Securely fixed when load
tested under approx. 75 kg
load (hang from it briefly
with two hands)
Effective brakes, securely
engaged brake
Locking in place, release
Accessories (e.g. patient
lifting pole, grab handle,
external safety sides)
Fixing, damage, suitability