13.6 Servicing Plan
Components marked with an asterisk (*) only apply for electric beds.
Components
Action
Interval
Information
All screw connec-
tions and/or metal
safety caps
Check that they are
firmly positioned; tight-
en and/or replace if
necessary.
At least once a year
Use suitable tools.
Connecting bolts on
the removable
headboard and
footboard
Clean and slightly
grease, if necessary.
At least once a year
Use resin-free and
acid-free grease
(e.g. grease C).
If machine washed, at
least after every 25
washes. Cleaning
Pivot and friction
bearings
Clean and lightly
spray.
As necessary
Use resin-free and
acid-free oil spray
(e.g. lubricant A).
Bowden cable for
CPR lowering of
backrest
Check that they are
correctly adjusted and
routed.
At least once a year
There must be no
sharp bends or
kinks!
Lead-acid batter-
ies*
Check whether
charged and ready for
use.
Recommendation: We
recommend that re-
chargeable batteries be
replaced after 5 years
(wear and tear part).
Connect bed to
mains supply.
Earth wires*
Check for tears and
tightness of screws.
At least once a year
Replace torn or
damaged earth
wires.
Damaged coating
Touch up any dam-
aged areas of the
coating:
As necessary
Special-purpose
paint (mechanically
stable anti-corrosive
protection) in a suit-
able colour.
O-rings (‘sealing
rings’) on the elec-
tric plugs*
Always check whether
they are present
and/or damaged.
Whenever plugs have
been disconnected (e.g.
after replacing actuators
or during troubleshoot-
ing)
The plugs of new
components are
always fitted with a
new O-ring.
Replace O-rings with original LINAK O-rings only. Check the corresponding sockets for dirt
or damage, and lubricate with Vaseline, before reinserting the plugs.
We recommend supplementing the specified maintenance work with appropriate preventive
measures to further enhance trouble-free and fail-safe operation of the bed. A guide to preventive
replacement of potentially relevant components can be obtained from our service centre: Contact
info: Service Address » 77
For motorised beds we recommend, as a guideline, that an annual DGUV A3 inspection be carried
out by our qualified Service Centre staff, with a certificate of adherence to the 2% error rate (see
also the DGUV A3 accident prevention regulations: § 5, Table 1B).