Troubleshooting
Troubleshooting for Electric Beds
96
Instruction Manual
Call Up or Delete Last Faults Saved
In addition, the last error code saved in the control unit can be called up and also deleted using the
handset.
Indication of error codes on the handset. If no error code can be called up, there was either no
fault or the memory was deleted.
15.1.2.2 Optional Equipment: Handset and Locking Box
Depending on the features which the bed incorpo-
rates, the bed has a locking box. This indicates the
existing fault using the LEDs on the buttons (see
picture).
Locking Box
Possible Faults
Solution
E
F
1
2
2
Í
Í
Thigh rest motor (socket 5)
limit switch
Check cable + plug
Check/replace motor
Í
Í
Handset fault
An internal fault in the control
unit
Replace handset with a new one
Replace control
Call Up or Delete Last Faults Saved
Display Last Fault
Delete Last Fault
Press both buttons at the same time
Press both buttons at the same time for 5 sec-
onds until the procedure is confirmed with a
beep.