What to do if my Stiegelmeyer Medical Equipment actuator runs briefly then stops?
- MMatthew TaylorSep 10, 2025
The drive may be overloaded. Remove the overload and retest. Also, check if one or more motors are not connected and connect all motors.
What to do if my Stiegelmeyer Medical Equipment actuator runs briefly then stops?
The drive may be overloaded. Remove the overload and retest. Also, check if one or more motors are not connected and connect all motors.
What does it mean if locking LEDs on handset/locking box alternate between orange and green on Stiegelmeyer Deka?
There is a serious problem with the control unit, and for safety reasons, all functions are locked. Unlock the control unit; if the fault occurs again, have the actuator system checked and inform your operator about any necessary repairs.
Why is operation with a sufficiently charged battery only possible for a short time on my Stiegelmeyer Deka?
The end of the lead-acid battery life has likely been reached. Exchange the lead-acid battery and inform your operator about any necessary repairs.
Manufacturer | Stiegelmeyer |
---|---|
Adjustable Height | Yes |
Knee Break Adjustment | Yes |
Side Rails | Yes |
Electrical Operation | Yes |
Wheels | Yes |
Defines target user groups and their responsibilities for safe bed operation.
Explains safety warning symbols and icon meanings used in the manual.
Essential safety rules for correct and safe operation of the hospital bed.
Precautions for mains cables, connections, and electrical components.
Guidelines on actuator limits and safe handset handling.
Warnings and measures to prevent crushing or entrapment during bed adjustments.
Information on fire prevention and electromagnetic compatibility requirements.
Defines the bed's intended use and lists conditions where it should not be used.
Detailed description of the bed's parts like chassis, actuators, and safety sides.
Key technical details: dimensions, weight, adjustments, and electrical data.
Essential safety guidelines and checks required before the bed's initial use.
Requirements for the operating environment and specific safety for electric beds.
Procedures and checks required when a new patient uses the bed.
Instructions for extending or shortening the mattress base.
Steps for removing and inserting headboard/footboard infill panels.
Procedures for safely moving, braking, and immobilising the bed.
Step-by-step guides for mechanical adjustments: backrest, thigh, leg, height, tilting.
Overview of electric components, control units, handset functions, and locking.
Instructions for raising and lowering swivelling safety sides.
How to activate the CPR position using handset or locking box.
Manual procedure for rapidly lowering the backrest in critical situations.
Safety guidelines, manual, and machine cleaning procedures.
Service addresses, safety during maintenance, and recommended servicing plans.
Safety info and guides for replacing batteries, handset, control unit.
Guide to identifying and rectifying faults in electric beds.
Guidelines for proper disposal of the bed, components, and gas springs.
Instructions for installing optional accessories like safety sides.
Includes inspection reports, user checklists, and declarations of conformity.