A-96.250.511 / 091120
AMI Turbiwell
Default Values
104
10. Default Values
Operation
Sensors Filter Time Const.: ........................................................................30 s
Hold after Cal.:............................................................................300 s
Alarm Relay .........................................................................same as in Installation
Relay 1 & 2 .........................................................................same as in Installation
Input .........................................................................same as in Installation
Logger Logger Interval:........................................................................ 30 min
Clear Logger:.................................................................................. no
Installation
Sensor Dimension:................................................................................... FNU
Flow; Flow measurement: ..........................................................None
Signal Output
1 & 2
Parameter:.......................................................................Meas. value
Current loop:........................................................................4 20 mA
Function:.....................................................................................linear
Scaling: Range low:...................................................... 0.0 FNU/NTU
Scaling: Range high: .................................................. 10.0 FNU/NTU
Alarm Relay Alarm high: .................................................................. 200 FNU/NTU
Alarm low:................................................................. 0.000 FNU/NTU
Hysteresis:..................................................................... 10 FNU/NTU
Delay: .............................................................................................5 s
Sample Flow: Flow Alarm: .............................................................yes
Sample Flow: Alarm High: ......................................................40.0 l/h
Sample Flow: Alarm Low: ......................................................... 6.0 l/h
Case temp. high: ....................................................................... 65 °C
Case temp. low:........................................................................... 0 °C
Relay 1 & 2 Function:.............................................................................limit upper
Parameter:.......................................................................Meas. value
Setpoint: ...................................................................... 100 FNU/NTU
Hysteresis:.................................................................. 5.00 FNU/NTU
Delay: .............................................................................................5 s
If Function = Control upw. or dnw:
Parameter: Meas. value
Settings: Actuator: Frequency
Settings: Pulse Frequency: .................................................120/min
Settings: Control Parameters: Setpoint:................... 100 FNU/NTU
Settings: Control Parameters: P-band: ................... 5.00 FNU/NTU
Settings: Control Parameters: Reset time:..................................0 s