A-96.250.511 / 091120
AMI Turbiwell
Program Overview
80
8. Program Overview
For explanations about each parameter of the menus see Program
List and Explanations, S. 86.
Menu 1 Messages: informs about pending errors and mainte-
nance tasks and shows the error history. Password protection
possible. No settings can be modified.
Menu 2 Diagnostics: is always accessible for everybody. No
password protection. No settings can be modified.
Menu 3 Maintenance: is for service: Calibration, simulation of
outputs and set time/date. Please protect with password.
Menu 4 Operation: is for the user, allowing to set limits, alarm
values, etc. The presetting is done in the menu Installation
(only for the System engineer). Please protect with password.
Menu 5 Installation: Defining assignment of all inputs and out-
puts, measuring parameters, interface, passwords, etc. Menu
for the system engineer. Password strongly recommended.
8.1. Messages (Main Menu 1)
Pending Errors Pending Errors 1.1.5* * Menu numbers
1.1*
Message List Number 1.2.1*
1.2* Date, Time
Maintenance List Maintenance List 1.3.5*
1.3*