AMI Turbiwell
Troubleshooting
76
7.4. Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is
exceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error (blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked E0xx (bold and orange)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked E0xx (bold and red)
Error or fatal Error
Error not yet acknowledged.
Check Pending Errors 1.1.5 and take
corrective action.
Navigate to <Messages>/
<Pending Errors>
Press [ENTER] to acknowledge the
Pending Errors.
The Error is reset and saved in
the Message List.
12 l/h
HOLD 14:10:45
R1
0.04 FNU
R2
1.1
Message List
Pending Errors
Maintenance List
Messages
1.1.5
Pending Errors
Error Code E002
Alarm low
<Enter> to Acknowledge