AMI Turbiwell
Program List and Explanations
88
2.4 I/O State
2.4.1- 2.4.2 Shows the actual status of all in- and outputs.
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
3 Maintenance
3.1 Verification
3.1.1 SWAN Verikit: Press [Enter] to select the Verikit you need for the ver-
ification. Further details see Verification, p. 64.
3.1.2 Other method: Other method is a wet verification with a standard
with a known turbidity filled into the measuring chamber instead of
the sample. Further details see Wet Verification, p. 72.
3.2 Matching
3.2.1 Solid Verikit
3.2.1.1 SWAN Verikit #: Press [Enter] to select the Verikt you want to match
or enter a new Verikit for matching. Further details see Matching, p.
64.
3.2.2 Liquid Verikit
3.2.2.1 SWAN Verikit #: Press [Enter] to select the Verikt you want to match
or enter a new Verikit for matching. Further details see Matching, p.
64.
3.3 Calibration
3.32.5 Calibration: After selecting Calibration press [Enter] and follow the in-
structions on the screen. Further details see Calibration, p. 59
Alarm Relay: Active or inactive
Relay 1 and 2: Active or inactive
Input: Open or closed
Signal Output 1 and 2: Actual current in mA
Signal Output 3: if option is installed
Available if FNU or NTU is chosen, see menu 5.1.2, p. 93.