49
o11
Dump controlDump
o12
Out of tolerance alarm output
14. Factory default (F8)
Navigating to menu [
F7
], then press and hold [ENTER] until menu [
F8
] displayed.
Lower window display list table.
Menu
Instruction
Operation instruction
F2 Int
F2 group
parameter
initialization
F2.1=1; buzer on
F2.2=1; enable tare operation
F2.3=20; manual zero range 20%( test mode is 4%)
F2.4=0; zero tracking is disabled
F2.5=3; dynamic checking range 3d
F2.6=5; filter coefficient 5
F2.7=0; disable power up zero
F3 Int
F3 parameter
initialization
F3. 1=0; BC360 continuous protocols
F3. 2=0; BC360 no checksum bits
F3. 3=2;9600 baud rate
F3. 4=0; 8 data bit / no checking bit
F3. 5=0; manual print without unit
F3. 6=1; node address
F6 Int
F6 parameter
initialization
F6.1=10 zero tolerance range is 1.0%F.S
F6.2=5 auto-spill correction frequency is 5
F6.3.1=1.0 start delay time sdt =1.0s
F6.3.2=0.5 comparison inhabit time cit =0.5s
F6.3.3=1.0 stable check time ttc =1.0s
F6.3.4=1.0 delay after dump over tdc =1.0s
F6.4=05 target tolerance checking frequency
F6.4A=1.0 target tolerance range 1.0%
F6.5=02 auto tare intervals
F6.6=0 auto feeding
F6.7=0 auto Dump
F6.8=4 material number
F6.9.1=2 material 1 2-speed feeding
F6.9.2=2 material 2 2-speed feeding
F6.9.3=2 material 3 2-speed feeding
F6.9.4=2 material 4 2-speed feeding
F6.9.10=0 abort last batching on powerup
F6.9.11=0 mixed batching
SH Int
Formula data
initialization
1-10 formula data
Target =1000target
Fine =100
slow add data
Preact =50 pre-act data
rECI PE =1 formula No.1
bAtcH =0
clear preset batch quantity