[PROGram]:LIST:DWELl <NRf+>,{<NRf+>}
Function: Specifies the time interval of each value (point) in a list to remain in effect.
The function accepts up to 100 parameters.
Parameters
0.001 to 129,600
Unit
S
Query
[PROGram]:LIST:DWELl?
Return
<NR2>,{<NR2>}
Example
LIST:DWEL .6,1.5,1.5,.4
NOTE:
A delay of ~100mSec is required after the [PROGram]:LIST:DWELl command if long sequences
are used, prior sending any additional command.
[PROGram]:LIST:VOLTage <NRf+>,{<NRf+>}
Function Specifies output voltage points in a list.
Voltage points are separated by commas.
The function accepts up to 100 parameters.
Parameters The range is limited by the Power Supply voltage programming range
(including UVP & OVP settings).
Unit
V
Query
[PROGram]:LIST:VOLTage?
Return
<NR2>,{<NR2>} 5-digits format
Example
LIST:VOLT 2.0,2.5,3.0; LIST:VOLT MAX,2.5,MIN;
NOTE:
A delay of ~100mSec is required after the [PROGram]:LIST:VOLTage command if long
sequences are used, prior sending any additional command.
[PROGram]:LOAD <NR1>
Function
Loads a sequence from a memory previously stored by the STOR command.
Parameters
1|2|3|4
Query
[PROGram]:LOAD?
Return <NR1> 1|2|3|4.
If any loaded sequence data has changed, but has not been stored yet, or
if no sequence is loaded, the reply is 0.