93
MODE?
Query Returns the Power Supply operation mode.
If Power Supply is off, returns OFF.
If Power Supply is on, returns CV for Constant Voltage, CC for Constant Current,
or CP for Constant Power
Return
<DSC> OFF|CV|CC|CP
MS?
Query Returns the Master/Slave setting
Single, Master + # (# - number of slaves units connected) or Slave
Return
<NR1>
NOTE:
Power Supply does not execute commands in the Slave mode.
5.10.7 Global Output Commands
General
Global commands are received by all Power Supplies connected to the BUS, without any need of
individual addressing commands.
All Power Supplies execute commands immediately. There is no acknowledgment back to the PC.
A delay of at-least 10mSec must be set after each global command. Error messages are not
reported back to the issuing PC.
GRCL <NR1>
Function
Refer to RCL command description.
Parameters
1,2,3,4
Returns
No reply
GRST
Function: Refer to RST command description.
Returns
No reply
GSAV <NR1>
Function
Refer to SAV command description.
Parameters
1,2,3,4
Returns
No reply
GPC <NR2>
Function
Refer to PC command description.
Returns
No reply