89
5.10.4 Identification Commands
IDN?
Query Returns the Power Supply model identification as an ASCII string (one comma,
no spaces)
Returns
<SRD>
Example
TDK-LAMBDA,G100-50
REV?
Query
Returns the software version as an ASCII string
Returns
<SRD>
Example
G:XX.XXX
SN?
Query
Returns the supply’s serial number. Up to 12 characters in any format
Returns
<SRD>
Example
111111-22222
DATE?
Query
Returns the date of last calibration. Format: yyyy/mm/dd
Returns
<SRD>
Example
2017/12/17
5.10.5 Initialization Commands
CLS
Function
Clear Status. Sets FEVE and SEVE registers to zeroes
Returns
<DSC> OK
ADR <NR1>
Function
Send ‘ADR’ and address (0 to 31) to open a communication channel.
Returns
<DSC> OK
RMT <DSC>
Function Set the Power Supply control source (front panel or communication) to local,
remote or LLO mode.
Local Enable front panel control.
Remote Disable front panel settings change (settings preview only).
LLO same as Remote + Disable front panel unlock function by front panel.
Change mode by communication or AC recycle.
Parameters
0|1|2 or LOC|REM|LLO
Query
RMT?
Returns
<DSC> LOC|REM|LLO