114
NOTE:
A delay of ~100mSec is required after the [PROGram]:WAVE:CURRent command if long
sequences are used, prior sending any additional command.
[PROGram]:WAVE:TIME <NRf+>,{<NRf+>}
Function: Specifies the time interval of each slope between 2 points of a WAVE.
The function defines up to 100 parameters.
0.001 to 129,600 seconds. Default: 0.001s
NOTE:
A delay of ~100mSec is required after the [PROGram]:WAVE:TIME command if long sequences
are used, prior sending any additional command.
[PROGram]:WAVE:VOLTage <NRf+>,{<NRf+>}
Function Specifies output voltage points in a wave.
Voltage points are separated by commas.
The function accepts up to 100 parameters.
Parameters The range is limited by the Power Supply voltage programming range
(including UVP & OVP settings).
<NR2>,{<NR2>} 5-digits format
WAVE:VOLT 2.0,2.5,3.0; WAVE:VOLT MAX,2.5,MIN. 5 digit format
NOTE:
A delay of ~100mSec is required after the [PROGram]:WAVE:VOLTage command if long
sequences are used, prior sending any additional command.