120
STATus:QUEStionable:CONDition?
Query Returns the value of the Questionable Condition Group Condition register,
which is a read-only register that holds the real-time conditional status of
the power supply.
Refer to STATus:QUEStionable[:EVENt]? command for the complete
registers list.
Returns
<NR1> 0…65535
Example
00136
STATus:QUEStionable:ENABle <NR1>
Function Sets the value of the Questionable Condition Group Enable register. This
register is a mask for enabling specific bits from the Condition register to
the Event register.
Refer to STATus:QUEStionable[:EVENt]? command for the complete list of
the registers that can be masked.
Parameters
0…65535
Query
STATus:QUEStionable:ENABle?
Returns
<NR1> 0…65535
4.14.10 SYSTem Subsystem
SYSTem[:COMMunicate]:ADDRess <NR1>
Function
Sets the power supply communication address.
Parameters
0…31
Query
SYSTem[:COMMunicate]:ADDRess?
Return
<NR1> 0…31
NOTE
Power supply addressing is lost after an address change.
SYSTem[:COMMunicate]:BAUDrate <DSC>
Function
Sets a serial communication Baud Rate (including multi-drop).
Parameters
9600|19200|38400|57600|115200 or 0|1|2|3|4
Query
SYSTem[:COMMunicate]:BAUDrate?
Return
<DSC> 9600, 19200, 38400, 57600, 115200
SYSTem[:COMMunicate]:BOOL[:STATe] <DSC>
Function Select query Return format of Boolean (<Bool>) commands.
DIGIT – Boolean queries are replied with 0|1.
TEXT Boolean queries are replied with OFF|ON.
Parameters
DIGIT|TEXT or 0|1
Query
SYSTem[:COMMunicate]:BOOL[:STATe]?
Return
<DSC> DIGIT, TEXT