7. Instrument Features
2004-11 Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5 7-3
Parameters Characteristics
Electrical
Power supply
Auto-sensing: 100-120 V / 220-240 V
50/60 Hz
Power consumption
250 VA
Main fuse
T 2.0 A / 250 V (slow blow)
Approvals
DIN EN 61010-1ff, UL 3101-1, C22.2 No. 1010.1
Physical
Outer dimensions Width: 812 mm
Height: 355 mm
Depth: 566 mm
31.7 inches
13.9 inches
22.1 inches
Weight approx. 60 kg
Environmental
Ambient temperature
Operation +15°C - +30°C +59°F - +86°F
Non-operation -20°C - +60°C -4°F - +140°F
Relative humidity
Operation 90 % non-condensing
Overvoltage category II
Pollution degree 2
Usage Commercial
Noise level < 75 dBA
Method of disposal Electronic waste
(infectious waste)