Removal and Installation Procedures
Procedures fo
r Internal Modules (Lower)
Front Connecto r Board
You need a screwdriver with a T-15 Torx tip (items 1 and 2).
Removal. To r
emove the Front Connector board, follow these steps: (See
Figure 4-19 on page 4-40.)
1. Remove the h
andle unit, snaps, cosmetic covers, front-trim u nit, and EMI
covers.
2. Orient the
instrument so that the top is on the work surface.
3. Disconnect the cables from J100 and J200 on the Front Connector board.
4. Remove the screw securing the Front Connector board to the chassis.
5. Grasp the board and pull it gently away from the chassis.
Installation. To install, reverse this procedure.
AWG7000B and AWG7000C Series Service Manual 4–39