-%!" !-"*-()! $!&.
!-"*-()! !-%"%/%*)
4Ć4
0)/%*)' !./. -!-!,0%.%/!.
1. Power on the oscilloscope and allow a 20 minute warmĆup before perĆ
forming this procedure; adjust the  control to display the
readout and the  control to display waveforms.
2. Disable the dual delay with the following menu selections.
a. Press the  button and select  from the main menu.
b. Select  until you can select 0' !'1 %.'! .
3. Press the   button and set  to "".
4. Install the probe on  . Connect the probe tip to   on
the frontĆpanel; connect the probe ground to the ground barrel of an
unused input BNC (see Figure 4Ć2).
Some functional checks require that you install the probe on connectors
other than  . All functional checks use the   on the
frontĆpanel as the signal source.
%#0-! 32!./ **&0+ "*- 0)/%*)' !./.
!-%"1 /$! -*! *(+!)./*- 0/+0/
1. Press the   button until the readout indicates that only
channel 1 is selected.
2. Press the  button.
3. Press the   button and set  to .
4. Set the volts/div scale to 1 V and vertically center the display.
5. Set the sec/div scale to 500 ms.
6. Press the  button and set D to ).
7. Align the active cursor to the top of the signal using the General Purpose
Knob.