#/$,/* +!# #010
  #/3'!#  +2 )
4Ć17
8. Connect the DC calibration generator to the   input as shown in
Figure 4Ć7.
DC Calibration Generator
Precision Cable
'%2/# 65/'%%#/ #3#) #01 #12-
9. Set the DC calibration generator for a 200 mV output.
10. Press the     button.
11. Check that the Trigger Level Accuracy readout is in the range of 169 mV
to 231 mV.
12. Press the   button and set  to  ))'+%.
13. Press the     button.
14. Check that the Trigger Level Accuracy readout is in the range of 169 mV
to 231 mV.
15. Press the   button, displaying the delay trigger
menu.
16. Set  to 2+0 $1#/ and  to &.
17. Press the     button.
18. Check that the Trigger Level Accuracy readout is in the range of 169 mV
to 231 mV.
19. Disconnect the test setup from the oscilloscope.
20. Press the   button and set  to '0'+%.
21. Press the   button and set  to $$.
&#!(  ,2-)'+%  +"4'"1&
.2'-*#+1 #.2'/#" One leveled sine wave generator (item 8), one preciĆ
sion coaxial cable (item 5), and one 50 W termination (item 3).
1. Display the channel to be verified, turning all others off.
2. Connect the output of the sine wave generator to the channel to be
verified as shown in Figure 4Ć8.