EasyManua.ls Logo

Tektronix TAS 465 - Check DC Coupling Bandwidth; Figure 4 7: Trigger Level Test Setup

Tektronix TAS 465
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
#/$,/* +!# #010
  #/3'!#  +2 )
4Ć17
8. Connect the DC calibration generator to the  input as shown in
Figure 4Ć7.
DC Calibration Generator
Precision Cable
'%2/# 65/'%%#/ #3#) #01 #12-
9. Set the DC calibration generator for a 200 mV output.
10. Press the     button.
11. Check that the Trigger Level Accuracy readout is in the range of 169 mV
to 231 mV.
12. Press the   button and set  to  ))'+%.
13. Press the     button.
14. Check that the Trigger Level Accuracy readout is in the range of 169 mV
to 231 mV.
15. Press the   button, displaying the delay trigger
menu.
16. Set  to 2+0 $1#/ and  to &.
17. Press the     button.
18. Check that the Trigger Level Accuracy readout is in the range of 169 mV
to 231 mV.
19. Disconnect the test setup from the oscilloscope.
20. Press the   button and set  to '0'+%.
21. Press the   button and set  to $$.
&#!(  ,2-)'+%  +"4'"1&
.2'-*#+1 #.2'/#" One leveled sine wave generator (item 8), one preciĆ
sion coaxial cable (item 5), and one 50 W termination (item 3).
1. Display the channel to be verified, turning all others off.
2. Connect the output of the sine wave generator to the channel to be
verified as shown in Figure 4Ć8.

Table of Contents

Related product manuals