EasyManua.ls Logo

Tektronix TAS 465 - Figure 5 3: Low Frequency Output Compensation Calibration

Tektronix TAS 465
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
!&0./(")/ -* "!0-".
  "-1% " )0'
5Ć7
Pulse Generator
Precision Cable
50 W Termination
DualĆinput
Coupler
%$0-" 54*2 -",0") 3 0/+0/ *(+")./%*) '%-/%*) "/0+
8. Set the pulse generator for fast rise period of 1 ms and a 4 division
display.
9. Press the     button.
10. Press the  button and set D to ).
11. Set the cursors 5 divisions apart using the General Purpose Knob and
the  button.
12. Set the volts/div scale to 20 mV.
13. Press the  button.
14. Set the channel 2 volts/div scale to 20 mV.
15. Press the   button and make the following selections
from the menu (channel 2):
H Set  to 
H Set  to ##
H Set  to ##
H Set  to 0''
16. Position the channel 2 trace to the center vertical graticule line.
17. Set  to .
18. Adjust R111 located on the Display Driver board for minimum vertical
movement of the readout (over the entire graticule area).
19. Disconnect the test setup from the oscilloscope.

Table of Contents

Related product manuals