EasyManua.ls Logo

Tektronix TAS 465 - Check X Axis Gain; Figure 4 9: X Axis Gain Test Setup

Tektronix TAS 465
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
*'*%& +,+
  *." &-$
4Ć19
9. Disconnect the test setup from the oscilloscope.
10. Repeat this procedure until all input channels are verified.
!# 1/"+ "&
)-"(%&, )-"* One pulse generator (item 10) and one precision
coaxial cable (item 5).
1. Display channel 1, turning all others off.
2. Connect the output of the pulse generator to the  input as shown
Figure 4Ć9.
Pulse Generator
Precision Cable
" -* 101/"+ "& +, ,-(
3. Set the output of the pulse calibration generator for 50 mV.
4. Press the  button.
5. Set the volts/div scale to 10 mV.
6. Center the display using the vertical  control.
7. Press the   button and set  to &.
8. Check that the amplitude of the XĆaxis signal is 4.8 to 5.2 divisions.
9. Set  to .
10. Disconnect the test setup from the oscilloscope.

Table of Contents

Related product manuals