Specifications (cont.)
Vertical
Offset range Scale range Offset range
g
1mV/divto9.95mV/div ±100 mV
10 mV/div to 99.5 mV/div ±1V
100 mV/div to 995 mV/div ±10 V
1 V/div to 10 V/div ±100 V
Offset Accuracy, all
ranges
± [0.005 ¢ | offset -- position | +0.1 div]
Note: Convert both the constant offset and position terms to volts
by multiplying by the volts/div setting.
Horizontal
Acquisition (horizontal)
resolution
Normal
(10,000 point record)
Fast trigger
(500 point record)
Maximum
acq rate, typical
TDS301xB,
TDS302xB
TDS303xB--
TDS306xB
TDS301xB,
TDS302xB
TDS303xB--
TDS306xB
400 wfms/s 700 wfms/s 2,600 wfms/s 3,600 wfms/s
Sample rate range
TDS301xB
TDS302xB,
TDS303xB
TDS304xB--
TDS306xB
Normal 100 S/s to
1.25 GS/s
100 S/s to
2.5 GS/s
100 S/s to
5GS/s
Fast trigger 5 S/s to
1.25 GS/s
5 S/s to
2.5 GS/s
5 S/s to
5GS/s
Seconds/division
range
4ns/divto
10 s/div
2ns/divto
10 s/div
1ns/divto
10 s/div