Appendix A: Specifications
TDS3000B Serie s User Manual
A-13
Specifications (cont.)
EMC certifications and compliances
EMC Compliance:
Australia/New Zealand
Meets the intent of Australian EMC Framework as demonstrated
to the following specification:
AS/NZS 2064.1/2
EMC Compliance:
Russia
This product was certified by the GOST ministry of Russia to be
in compliance with all applicable EMC regulations.
FCC Compliance:
U.S.A.
Emissions comply with FCC Code of Federal Regulations 47,
Part 15, Subpart B, Class A Limits
1
Emissions that exceed the levels required by this standard may occur when t his
equipment is connected to a test object.
2
To ensure compliance to the standards li sted above, at tach only high quality
shielded cables to thi s instrument. Hi gh quality shielded cables typically are braid
and foil types that have low impedance connections to shielded connectors at
both ends.
3
The increase in trace noise while subjected to a test field (3 V/m over the
frequency range 80 MHz to 1 GHz, wi th 80% am plitude modulation at 1 kHz) is
not to exceed 8 major divisions peak-to-peak. Ambient conducted fields may
induce triggering when the trigger threshold is offset less than 4 major divisions
from ground reference.
4
The increase in trace noise while subjected to a test field (3 V/m over the
frequency range 150 kHz to 80 MHz, wi th 80% amplitude modulation at
1 kHz) is not to exceed 2 major divisions peak-to-peak. Ambient conducted fi elds
may induce triggering when the trigger threshold is offset less than
1 major divisions from ground reference.